Bird-technologies 6091P Instrukcja Użytkownika Strona 56

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 55
Bird Model 6091 Calorimeter
44
Status Words (Ux) (Continued)
Function
Revision History Word (U2) - Contents include current six bytes contained in
writeable store, hardware and software revision levels and the revision level of
the IEEE-488 standard being implemented by the Calorimeter for GPIB
programming.
If Calorimeter requires repair, supply hardware and software revision level to ser-
vice center.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
-6091-XXXX X X Y Y Z Z 7 8 A A CRLF
Terminators
Address
IEEE-488 1978
Hardware Revision
Software Revision
Writeable Store
Header
Remarks
Status word returned only once for each Ux.
Machine Status-A19character string containing the current command settings.
Error Status-A17character string that indicates valid/invalid commands and
command options as well as results from the self-test (J0).
Revision History-A20character string containing hardware and software revi-
sion levels, writeable store data and the current instrument address.
Statements
National GPIB-PCII HP 200/300
Let CMD$="U1"
CALL IBWRT (UD%,CMD$)
OUTPUT 724; "U1"
Przeglądanie stron 55