Bird-technologies 5000-EX Instrukcja Użytkownika Strona 42

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 41
Bird Model 5000-EX Digital Power Meter
30
Bird 5011 and 5011-EF Terminating Sensors
General Corrected terminated average
power
Sensor Type Diode based terminated true
average power measurement
Frequency Range
5011
5011-EF
40 MHz to 4 GHz
40 MHz to 12 GHz
RF Power Range 10
µW to 10 mW
(–20 dBm to +10 dBm)
Max Power (Damage
Level)
2 W avg.
125 W peak for 5 µs
Peak/Average Ratio 12 dB max
Uncertainty
*
±(5% of reading ± 1 µW) (95% c.l.)
(excluding mismatch uncertainty)
(with correction factors for -EF)
RF Connector
5011
5011-EF
N Male
Precision N Male
Impedance, Nominal 50 ohms
Input VSWR:
5011
Typical
Maximum
5011-EF
Typical
Maximum
1.03 (36.6 dB return loss)
1.20 (20.8 dB return loss)
1.05 (32.0 dB return loss)
1.25 (19.1 dB return loss)
Power Supply From host instrument via cable